Kestra published ACE-PAS, the post-approval study FDA required as a condition of its ASSURE wearable cardioverter defibrillator approval, in the Journal of Cardiovascular Electrophysiology. The prospective single-arm registry fitted 21,612 patients between November 2021 and July 2025 across 35 states. All 303 analyzable ventricular tachycardia or fibrillation shock events converted with one or more shocks, meeting the primary effectiveness endpoint.

Median wear time was 23.1 hours a day, and about 30 percent of patients kept the device past 90 days at the same median. The inappropriate shock rate came in at 0.0065 per patient-month, below the prespecified safety goal. Ninety-four percent of patients had no false positive shock alarms.

Arrhythmia events were concentrated early. Some 2.6 percent of patients had at least one VT or VF event, and roughly 55 percent of episodes landed in the first month. Among patients with an ejection fraction of 35 percent or less, 90-day event rates ran 3.2 percent after myocardial infarction and 2.8 percent in ischemic cardiomyopathy, with 2.5 percent in nonischemic cardiomyopathy.

Kestra calls this the largest prospective real-world study of a wearable defibrillator. For anyone building a PMA device with a post-approval condition attached, the wear-time number is the one payers will read first, and it took almost four years of enrollment to produce.