FDA posted an early alert on November 10 covering 21 Arrow International products whose spring wire guide components carry a hydrophilic coating that may swell unevenly when hydrated with saline. The coating can clump and detach. Released particles can cause hydrophilic polymer emboli, with inflammation, thrombosis, ischemia, tissue injury, stroke or death among the listed harms.
The list includes Arrow Marked Guidewires in hydrophilic coated nitinol and stainless steel, one, two and three lumen Arrowg+ard Blue Advance PICCs and JACCs with 80 cm or 130 cm wires, pressure injectable PICCs, and the Arrow VPS Precision Stylet plus the PICC kits preloaded with it. Arrow, a Teleflex subsidiary, sent its customer letter on November 5 and reported no serious injuries or deaths tied to the issue as of that date.
FDA tells users to check inventory, stop using and distributing affected product, quarantine it, and return it through Teleflex customer service. Early alerts go out before a recall classification, so the class designation is still ahead. For anyone shipping a hydrophilic coated wire, the failure mode here is the one worth reading twice: not the coating chemistry, the hydration step a nurse performs at the bedside.