FDA posted an early alert on November 10 covering 16 reference numbers of Intersurgical 22mm Smoothbore breathing tubes and systems. Intersurgical wrote to affected US customers on November 3 telling them to stop use immediately, quarantine stock and return the product.
The defect sits in the bonding step. Intersurgical uses tetrahydrofuran solvent to bond soft connectors onto the Smoothbore tube. Reports from markets outside the US described connectors coming off, and the company's testing found some units were built with too little solvent for a secure bond. A connector that detaches during use can interrupt ventilation until someone restarts it by hand, and FDA's notice ties that to hypoxia, organ failure, brain damage or death depending on the patient and the delay.
Affected codes include REF 5018000, 5025000, 5025310, 5026000, 5183000, 5191000, 5191850, 5194850, 5804000, 5805000 and 5805030, with lot lists running to 18 lots on 5805030 alone. Intersurgical reported no serious injuries or deaths as of November 3. The same day FDA posted this alert, it posted one for Arrow International spring wire guide components.
A solvent volume that drifts low enough to fail a bond is a process control problem, not a design one. If your product has a solvent or adhesive joint, the useful question is whether your release testing would catch an underdosed lot or only a grossly failed one.